Samba 200 firmware update description. Version 1.0.7 (for Samba 201 and Samba 202) New features: Bug fixes: Zero baseline function could fail when tying to zero a low pressure transducer when atmosphere pressure is very high. This update require the use of the unpgraded 1ch analog cable descibed in 1.0.6. Remaining issues: Non recorded. MDN 0805022 ********************************************************* Version 1.0.6 (for Samba 201 and Samba 202) New features: Bug fixes: Removed problem with spikes on the analog out port. Version 1.0.4 did not solve the problem on all control units. This update require the use of an unpgraded 1ch analog cable. Remaining issues: Non recorded. -- Do not upgrade before receiving the new analog cable (only 1 ch units) -- MDN 0805013 ********************************************************* Version 1.0.4 (for Samba 201 and Samba 202) New features: Bug fixes: Removed problem with spikes on the analog out port. Version 1.0.3 could cause this on some control units. Remaining issues: Non recorded. -- All users should upgrade -- MDN 080423 ********************************************************** Version 1.0.3 (for samba 201 and Samba 202) New features: Support for high pressure transducers. 5, 10, and 17 bar pressure range supported Bug fixes: Unit sometimes refusing to start after running battery empty, even after battery were recharged and power adapter connected. <> To prevent restart at criticaly low battery voltage a shutdown feature were added based on battery voltage. Remaining issues: Non recorded. -- All users should upgrade -- MDN 080220