Samba Resp 420 transducer

Samba Resp 420 is a fiberoptic pressure transducer for clinical measurements of the pressure in the lungs and airways.

Samba Resp 420 transducer

The regulatory approval allows for insertion through the trachea. Due to the small transducer size it's particularly valuable for measurement of the airway pressure on neonates.
The product is CE marked as a Class IIb product (permission to use on human patients within EU only).

The Samba Resp 420 is currently available in a version compatible with the Samba 3200 Control Unit.


The Samba Resp 420 catheter is not FDA approved.

Product specifications

Sensor diameter 0.42 mm (1.3 F)
Fiber diameter 0.25 mm (0.75 F), incl. standard coating
0.40 mm (1.2 F), incl. radiopaque coating
Calibration Factory calibration (on-site calibration not needed)
Tolerated bend radius 10 mm (0.4 in)
Pressure range -50 to 350 cm H2O
Accuracy ±0.5 mbar and ±2.5% of reading (-50 to 250 cm H2O)
or ±4% of reading (250 to 350 cm H2O)
Temperature coefficient <0.2 cm water/ºC (20-45ºC/ 68-113ºF)
Long term stability <0.5% of range
Storage temperature -40 to +80ºC (-40 to 176ºF)
Total transducer length 4 m (13,1 ft)
Bare fiber length 14.5 cm (5.7 in)
Length of loose PTFE tubing 20 cm (7.9 in), or
60 cm (23.6 in)
Coating Standard, or
Radiopaque (visible under X-ray)
Regulatory approval CE, class IIb
Compatible Samba system Samba 3200 Control Unit